Oxigraf Laser Diode Oxygen Gas Analyzer
Specifications for the Model O2
 Oxygen Gas Analyzer
 Home
 Product Line
Applications
 Projects
 Tech Support
 Corporate

Model O2 Oxygen Gas Analyzer Specifications

   

PERFORMANCE CONDITIONS

The performance specifications for the Model 02 oxygen gas analyzer are valid under the following conditions:

Ambient temperature

5 deg to 40 deg C operating;
 -20 deg to 60 deg storage

Gas inlet temperature

0 deg to 45 deg C

Cell pressure

70 to 115 kPascal (500 to 900 mm Hg)

Warm-up for full accuracy

5 minutes

Altitude

Two point recal desirable after change in altitude of 2,000 feet

Humidity

0 to 95%, non-condensing

     

PERFORMANCE SPECIFICATIONS

XC MODE

LN MODE

Range

5-100%

5-100%

Accuracy

+- 0.5%

+- 0.2%

Short term stability

+- 0.1%

+- 0.03%

Cross sensitivity

+- 0.2%

+- 0.2% [ NOTE]


Response time

120 ms at 200 mL/min flow rate measured at inlet; 12 software averaging filters available from front panel

Display resolution

0.1 % front panel; 0.02% serial outputand PC display

Software

DOS Plot and Graph functions provided; also export to spreadsheets

NOTE : Limited to air mixtures and CO2 concentrations under 6%.

   

ELECTRICAL SPECIFICATIONS

 

Sensor power requirements

12 +-  0.6 VDC, 2.5 A max, 0.8 A typical

External power supply

95 to 250 VAC; 47 to 63 Hz

Analog output

0 to 1.0 volts for 0 to 100% oxygen, 1.00K
+- 1% output resistance

Digital output

RS232:  9600 baud default, 8 bit, no parity

Rule
Contact us!
[Home] [Product Line] [Applications] [Projects] [Tech Support] [Corporate] [Contact Us]

1170 Terra Bella Ave   Mountain View CA 94043   tel 650.237.0155  fax 650.237.0159  oxigraf@oxigraf.com
© 2001 Oxigraf Inc.  All rights reserved.