Model O2 Oxygen Gas Analyzer Specifications |
|
|
|
| |
|
|
PERFORMANCE CONDITIONS |
The performance specifications for the Model 02 oxygen gas analyzer are valid under the following conditions: |
|
Ambient temperature |
5 deg to 40 deg C operating; -20 deg to 60 deg storage |
|
Gas inlet temperature |
0 deg to 45 deg C |
|
Cell pressure |
70 to 115 kPascal (500 to 900 mm Hg) |
|
Warm-up for full accuracy |
5 minutes |
|
Altitude |
Two point recal desirable after change in altitude of 2,000 feet |
|
Humidity |
0 to 95%, non-condensing |
|
|
|
|
|
|
| |
|
|
|
PERFORMANCE SPECIFICATIONS |
XC MODE |
LN MODE |
|
Range |
5-100% |
5-100% |
|
Accuracy |
+- 0.5% |
+- 0.2% |
|
Short term stability |
+- 0.1% |
+- 0.03% |
|
Cross sensitivity |
+- 0.2% |
+- 0.2% [
NOTE] |
|
|
|
|
|
|
Response time |
120 ms at 200 mL/min flow rate measured at inlet; 12 software averaging filters available from front panel |
|
Display resolution |
0.1 % front panel; 0.02% serial outputand PC display |
|
Software |
DOS Plot and Graph functions provided; also export to spreadsheets |
|
|
|
|
NOTE : Limited to air mixtures and CO2 concentrations under 6%. |
|
|
| |
|
|
ELECTRICAL SPECIFICATIONS |
|
|
Sensor power requirements |
12 +- 0.6 VDC, 2.5 A max, 0.8 A typical |
|
External power supply |
95 to 250 VAC; 47 to 63 Hz |
|
Analog output |
0 to 1.0 volts for 0 to 100% oxygen, 1.00K +- 1% output resistance |
|
Digital output |
RS232: 9600 baud default, 8 bit, no parity |
|
|
|
|
 |
|
|
 |
|
[Industrial Applications]
[Oxygen Depletion Monitor]
[Cardio Applications]
[Home]
[Oxygen Analyzers and Sensors]
[Applications]
[Projects]
[Tech Support]
[Corporate]
[Contact Us]
|
|
|
|
|
1170 Terra Bella Ave Mountain View CA 94043 tel 650.237.0155 fax 650.237.0159
oxigraf@oxigraf.com © 2004 Oxigraf Inc. All rights reserved. |
|
|
|
|
OXIGRAF |
|
|
|
|
|
|